Fluorescence photometer PHOT and wavelength changer HSWES
fluorescence photometer PHOT

The fluorescence exciting light is fed into the photometer via a light guide. The fluorescence light is detected by a photo-multiplier system.

A slit allows both, restricting the exciting light to the preparation and detecting the fluorescence light exclusively from the excited area.

A light intensity monitor detects the exciting light intensity. Dividing the fluorescence light signal by the monitor signal deprives fluctuations in the exciting light intensity.

 



wavelength changer

The ultra-fast wavelength changer HSWES changes the fluorescence exciting light from one wavelength wave length to the other within 75µs.

A signal sorter unit sorts the fluorescence signal due to the two exciting wavelengths into two DC channels.

 
Design & Implementation: Lucy Kuijk.
Copyright © 2012 Scientific Instruments Heidelberg. All Rights Reserved.  Joomla! is Free Software released under the GNU/GPL License. Valid XHTML and CSS